How important is the Spectacle Frame Bridge Deformation Tester and how to use it?

A pair of glasses is like a pair of regenerated eyes for those of us with poor eyesight, 
which brings great convenience to our lives, so the stability of the use of the glasses
bridge becomes particularly important. The function of the glasses bridge deformation 
tester is The measurement frame has a certain ability to retain the lens, which ensures 
that the frame can be safely worn for more than 2 years, and prevents the lens from 
falling off due to slight impact, which affects the wearing effect. It is used by mounting 
the glasses horizontally on the test unit with the outer lens facing down. One lens is 
clamped in place while a vertical force is applied to the other lens. After the force 
is removed, measure the relative horizontal permanent deformation of the side under 
test and perform a compliance assessment.


Spectacle Frame Bridge Deformation Tester

Product Introduction
Reference standards: EN1836:2005, ISO12870:2004, GB/T14214:2003, etc.
Test reason: The measurement frame has a certain lens retention ability, ensuring 

that the frame can be safely worn for 2 years or more, preventing the lens from 

falling off due to a slight impact, affecting the wearing effect.

Parameters 
1. Pneumatic device: 30 Pa
2. Maximum pressure: 5N
3. Electronic display and automatic timing device
4. Size: 30 × 32 × 33 CM (length × width × height)
5. Weight: 16 Kg

Standards 
EN1836:2005, ISO12870:2004, GB/T14214:2003, etc.

Instrument installation
1. Test environment: temperature 23 ± 2 ° C, relative humidity 60 ± 2%.
2. Place the instrument on a relatively flat test platform.
3. Power connection: AC220V, 50Hz, 5A.