This equipment is used to measure the impact resistance of metal materials under dynamic load, so as to judge the properties of materials under dynamic load. The absorbed energy of the sample is obtained by the way the difference between the potential energy before the pendulum impact and the residual potential energy after the impact is displayed on the dial. The maximum impact energy is 300J, and a 150J pendulum is attached, and the section of the sample used is (10×10) mm. This machine has large impact energy and is suitable for ferrous metals with high impact toughness, such as steel and its alloys.
The electric appliance of this machine controls automatic swinging, impacting and swinging, which is easy to operate and has high work efficiency. After the sample is broken, the remaining energy is used to automatically lift the pendulum to prepare for the next impact test. Therefore, it can better reflect its advantages in the laboratory that continuously performs impact tests and the metallurgical and machinery factories that perform a large number of impact tests.
The main body of the machine is made of cast steel, which makes it rigid and stable during the whole test process, with strong impact resistance, not easy to deform, accurate test results and long service life. The use of high torque clutch and high power motor makes it more stable and vibration-free when taking and placing the pendulum.
This machine complies with JJG 145-2007 "Verification Regulations of Pendulum Impact Testing Machine".
It can meet the requirements of international standards such as American Standard, European Standard ASTM E23, EN 10045, ISO 148, ISO 83 and so on.
1. Impact energy: 300J 150J
2. Dial scale range and division value: Energy range: 0-300J 0-150J Graduation value per small division: 2J 1J
3. Pendulum moment: 0-300J M=160.7695N.m 0-150J M=80.3848N.m
4. Pendulum pre-lifting angle: 150°
5. Distance from the rotation center of the pendulum axis to the impact point (sample center): 750mm
6. Impact speed: 5.2m/s
7. Span of sample support: 40mm
8. Fillet of support jaws: R (1.0~1.5) mm (1mm is for special order)
9. Blade curvature radius: R (2.0~2.5) mm (8mm is for special order)
10. Inclination angle of the support surface of the sample support: 11°
11. The included angle of the impact blade: 30°
12. Thickness of impact blade: 16mm
13. Specimen size: 10×10 (7.5 or 5)×55 mm
14. Dimensions of the host (length × width × height): (2150 × 640 × 1400) mm
15. Net weight: about 420Kg
16. Power supply: three-phase four-wire system 380V 50Hz 400W Main motor 250W
17. Working conditions:
(1) The room temperature is within the range of 20°C±5°C;
(2) The relative humidity is not more than 85%;
(3) In an environment with no corrosive medium, no vibration, and no strong electromagnetic field interference around;
(4) Installed on a concrete foundation with a thickness of not less than 200mm or fixed on a foundation greater than 1400Kg;
(5) Adjust the levelness of the installation reference surface on the base to within 2/1000.
1. 1pc 300J main engine
2. 1pc 150J pendulum and 1pc 300J pendulum
3. 1pc electric control box
4. 1pc Span to find the positive board
5. One sample centering device
6. One disassembler (for pendulum change)
7. Four anchor bolts
8. A set of safety protection devices